Friday, June 14, 2013
Cross-Validation of IFN-γ Elispot Assay for Measuring Alloreactive Memory/Effector T Cell Responses in Renal Transplant Recipients
Assessment of donor-specific alloreactive memory/effector T cell responses using an IFN-γ Elispot assay has been suggested to be a novel immune-monitoring tool for evaluating the cellular immune risk in renal transplantation. Here, we report the cross-validation data of the IFN-γ Elispot assay performed within different European laboratories taking part of the EU RISET consortium. For this purpose, development of a standard operating procedure (SOP), comparisons of lectures of IFN-γ plates assessing intra- and interlaboratory assay variability of allogeneic or peptide stimuli in both healthy and kidney transplant individuals have been the main objectives. We show that the use of a same SOP and count-settings of the Elispot bioreader allow low coefficient variation between laboratories. Frozen and shipped samples display slightly lower detectable IFN-γ frequencies than fresh samples. Importantly, a close correlation between different laboratories is obtained when measuring high frequencies of antigen-specific primed/memory T cell alloresponses. Interestingly, significant high donor-specific alloreactive T cell responses can be similarly detected among different laboratories in kidney transplant patients displaying histological patterns of acute T cell mediated rejection. In conclusion, assessment of circulating alloreactive memory/effector T cells using an INF-γ Elispot assay can be accurately achieved using the same SOP, Elispot bioreader and experienced technicians in kidney transplantation (read more)
Print this post
Subscribe to:
Post Comments (Atom)
No comments: