Saturday, April 30, 2016
A reliable method for avoiding false negative results with Luminex single antigen beads; evidence of the prozone effect.
Luminex single antigen bead (SAB) assays have become an essential tool in monitoring the status of antibody to the Human Leucocyte Antigen (HLA) of patients both before and after transplantation. In addition SAB data is used to aid risk stratification to assess immunological risk of humoral rejection in solid organ transplantation (CTAG/BTAG guidelines). Increasingly laboratories are reporting false negative results at high antibody titre due to a prozone effect. Here we report a case study where the prozone effect led to a false negative antibody result that could have resulted in adverse outcome. We describe a method to reliably remove the prozone effect through heat inactivation and the addition of Ethylenediaminetetraacetic acid (EDTA) to the Luminex wash buffer (read more)
Print this post
Subscribe to:
Post Comments (Atom)
I was diagnosed as HEPATITIS B carrier in 2013 with fibrosis of the
ReplyDeleteliver already present. I started on antiviral medications which
reduced the viral load initially. After a couple of years the virus
became resistant. I started on HEPATITIS B Herbal treatment from
ULTIMATE LIFE CLINIC (www.ultimatelifeclinic.com) in March, 2020. Their
treatment totally reversed the virus. I did another blood test after
the 6 months long treatment and tested negative to the virus. Amazing
treatment! This treatment is a breakthrough for all HBV carriers.