Methods: Forty-five sera of highly immunized patients have been assessed by in-house modified C1q- and C4d-Luminex assays and compared with standard CDC and IgG-Luminex.
Results: Antibody specificities assigned by the C1q- and C4d-Luminex assay revealed an excellent concordance of 94% and 97% for HLA class I and II, respectively. Complement-fixing HLA class II antibodies were found less frequently among IgG antibodies compared with class I. Both C1q- and C4d-Luminex detected, on average, three times more specificities than CDC. Although we found a high correlation of mean fluorescence intensity values between C1q- and C4d-Luminex assays, IgG mean fluorescence intensity was not a suitable surrogate marker for the prediction of complement binding.
Conclusions: C1q- and C4d-Luminex assays are characterized by an increased sensitivity and specificity compared with CDC, the current standard in detecting complement-fixing HLA antibodies. Pretransplantation risk assessment for transplantation but also posttransplantation monitoring are important applications for both assays to improve overall allograft survival (read more) Print this post